6
Maintenance
Adding Calibration Compound
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
119
7. Remove the syringe from the calibration compound vial.
Figure 61.
Calibration compound vial
8. Inspect the 1/4 in. ferrule for damage. Replace it if necessary (P/N 95001-20310).
9. Reinstall the calibration compound vial (P/N 96000-40013) into the module.
a. Hold the vial vertically and place the nut on the neck of the vial.
b. Place the 1/4 in. ferrule over the neck of the vial and into the nut.
c.
Connect the vial to the calibration valve fitting with the nut and ferrule.
10. Reinstall the cover to the calibration compound vial compartment.
11. Position the GC next to the mass spectrometer.
12. Restart your system. See
“Starting Up the System after a Complete Shutdown”
Calibration
compound vial
Ferrule
Nut