6
Maintenance
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
83
Maintenance
This chapter describes routine maintenance procedures that you must perform to ensure
optimum performance of the TSQ Quantum XLS or TSQ Quantum GC mass spectrometer.
Optimum performance of the mass spectrometer depends on the maintenance of all parts of
the instrument.
You are responsible for maintaining your instrument properly by
performing the maintenance procedures on a regular basis
.
lists routine and infrequent mass spectrometer maintenance procedures. For
instructions on maintaining the GC or autosampler, refer to the manuals that come with the
GC or autosampler.
Note
When you perform maintenance procedures, be methodical; always wear clean,
lint-free gloves when handling the components of the ion source; always place the
components on a clean, lint-free surface; and never overtighten a screw or use excessive
force.
Table 6.
Maintenance procedures (Sheet 1 of 2)
Mass spectrometer
component
Procedure
Frequency
Procedure
location
Ion source
Cleaning the ion
volume
As needed*
Ion source
Cleaning the lenses
As needed*
(TSQ
Quantum XLS),
(TSQ
Quantum GC)
Ion source
Cleaning the entire ion
source
As needed*
(TSQ
Quantum XLS),
(TSQ
Quantum GC)
Ion source
Replacing the filament If the filament fails
(TSQ
Quantum XLS),
(TSQ
Quantum GC)
Inlet valve
Replacing the ball valve
seal
If there is an air
leak in the inlet
valve
Calibration compound Adding calibration
compound
If EZ Tune error
message indicates
low intensity of
calibration gas ions