Maintenance
Cleaning Ion Source Components
90
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
4. Undo the two latches that secure the lid of the ion source vacuum manifold, and open the
lid.
5. Disconnect the connectors from the lead pins at the EI/CI Source PCB, and L1, L2, L3
lens assembly.
Figure 48.
TSQ Quantum XLS ion source, installed
6. While holding the ion source assembly, loosen the two thumbscrews that secure the ion
source assembly to the Q0 quadrupole mount.
7. Pull the ion source assembly away from the Q0 quadrupole mount and 2 mm to the right
to clear the transfer line bellows.
8. Slide the ion source off the magnet yoke.
Q0 quadrupole mount
Lens L1, L2, L3 connector
EI/CI Source PCB
connector
Transfer line
Thumbscrew
Thumbscrew
CAUTION
Burn Hazard
. The ion source might be hot. Allow the ion source to cool
to room temperature before touching it.
Note
Wear clean, lint- and powder- free gloves when you handle the ion source.