6
Maintenance
Disassembling and Reassembling the TSQ Quantum XLS Ion Source Completely
108
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2. Shut down and vent the TSQ Quantum XLS. See
3. Wearing clean, lint- and powder- free gloves, remove the ion source. See
4. Remove the ion source from the magnet yoke.
Figure 58.
TSQ Quantum XLS ion source, disassembled
5. Pull the Q00 quadrupole lens assembly and the L1, L2, L3 lens assembly off the ion
source.
6. Pull the heater ring off the ion source.
7. Unscrew and remove the three base studs.
CAUTION
Shock Hazard
. Unplug the TSQ Quantum XLS mass spectrometer before
proceeding.
Filament retainer clip
Filament
Ion source block
L1, L2, L3 lens
assembly
Heater ring
EI/CI Source PCB
Base studs
Centering ring
Q00 quadrupole
lens assembly
Threaded standoff
Heater ring and lens
assemblies retainer clip