Changing Ionization Modes
Installing the Ion Volume
80
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Installing the Ion Volume
You install an ion volume after cleaning or to change ionization modes.
Y
To install the ion volume
1. Place the clean ion volume on the I/R tool with the arrows aligned. See
.
2. Turn the I/R tool handle to the lock position
.
3. Install the guide bar.
a. With the guide ball track facing left, insert the guide bar into the entry housing (see
b. Push the guide bar in as far as it will go, and then rotate it 90° clockwise to lock the
guide bar in the entry housing.
4. Prepare the inlet valve for insertion.
a. Make sure the inlet valve is closed.
shows the inlet valve lever is down for
closed.
b. Loosen the inlet valve knob counter-clockwise and remove the inlet valve plug.
Remove the inlet valve plug. The inlet valve plug prevents air from entering the
vacuum manifold in case the inlet valve is inadvertently opened.
5. Choose
Start > All Programs > Thermo Instruments > TSQ > TSQ Tune
to open the
EZ Tune window.
6. Click the
Probe
button in the EZ Tune toolbar. The Insert Probe message
appears. See
.
7. Inset the I/R tool into the inlet valve.
a. Insert the guide ball into the guide ball hole.
b. Slide the I/R tool forward into the inlet valve until the guide ball is at the guide bar’s
first stop. See
.
c.
Tighten the inlet valve knob clockwise to ensure a leak-tight seal.
8.
Click
OK
. The forepump evacuates the inlet valve. Wait for the safe to insert probe
message to appear (
); then click
OK
.
9. Once evacuation is complete, pull the inlet valve lever up to open the inlet valve.
Note
Wear clean, lint- and powder- free gloves when you handle a clean ion volume.
CAUTION
Make sure the arrows on the I/R tool and ion volume are aligned to avoid
damage to the ion source.