5
Changing Ionization Modes
Removing the Ion Volume
78
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
8. Remove the ion volume.
a. Slide the I/R tool into the vacuum manifold until the tip of the I/R tool is fully
inserted into the ion volume holder. The head of the arrow on the I/R tool
(
), when viewed through the window of the ion source manifold, should
not be visible.
Figure 43.
I/R tool inserted into the inlet valve
b. Turn the I/R tool handle counterclockwise to put the I/R tool into lock position
. Listen for a click, which indicates that the handle is fully engaged in
the lock position and is holding the ion volume.
c.
Withdraw the I/R tool (with the ion volume attached) until the guide ball reaches the
first stop (see
).
d. Pull the lever down to close the inlet valve.