4
Tuning and Calibrating
Displaying the FC-43 Mass Spectrum
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
65
• First mass:
50
• Last mass:
650
• Scan time:
0.500
Figure 31.
Define Scan dialog box
5. Click
Apply
to apply the settings.
6. Click the
Calibration Gas
button in the EZ Tune toolbar to turn on the flow of
calibration gas into the ion source.
7. Click the
Filament
button in the EZ Tune toolbar to turn on the filament.
8. Choose
Control > On
to start the mass spectrometer scanning.
9. Observe the mass spectrum. If your mass spectrum looks very different from the one in
“Diagnostics and Troubleshooting”
.......... ..Cal Gas Off .........On
.............Filament Off. . ....On