Maintenance
Disassembling and Reassembling the TSQ Quantum GC Ion Source Completely
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
111
Figure 59.
Disassembling the ion source
5. Pull the lens assembly off the ion source.
6. Pull the heater ring off the ion source.
7. Unscrew and remove the three base studs.
8. Remove the EI/CI Source PCB.
a. Place the ion source on a clean surface with the cartridge heaters down and the EI/CI
Source PCB up.
b. Press down on the ion source block to separate it from the EI/CI Source PCB.
9. Remove the retainer clip that secures the filament and centering ring to the ion source
block.
10. Remove the filament and centering ring from the ion source block.
Filament retainer clip
Filament
Ion source block
Lens assembly
Heater ring
EI/CI Source PCB
Base studs
Centering ring
Threaded standoff
Heater ring and
lens assembly
retainer clip
CAUTION
Shock Hazard
. Do not
pull
the EI/CI Source PCB out of the ion source
block. This might damage the EI/CI Source PCB.