3
Maintenance
Lamp Maintenance
42
Surveyor UV/Vis Plus Detector Hardware Manual
Thermo Scientific
Removing the Tungsten Lamp
To remove the tungsten lamp
1. Turn the detector off, unplug the power cord from the power receptacle on the back
panel, and allow sufficient time for the lamps to cool before you remove the lamp cover.
2. Open the front doors of the detector.
3. Using the 1/8 in. ball-end hex screwdriver (supplied), unscrew the captive retaining screw
on the lamp cover (see
4. Pull the cover straight back to expose the lamp assemblies.
5. Gently unplug the tungsten lamp lead from the detector (see
).
Take care not to twist the connector.
Figure 33.
Tungsten lamp assembly
6. Loosen and remove the thumbscrew that holds the lamp assembly in place.
7. Pull the assembly straight out and remove it from the detector.
CAUTION
Because intense UV light can damage your eyes, the detector automatically
turns off the lamps when you remove the lamp cover. As a secondary precaution, turn
off the detector and unplug it from line power.
CAUTION
Because the lamps run hot, turn off the power and allow sufficient time
(approximately 30 minutes) for the lamps to cool before you remove the lamp cover.
Tungsten lamp
connector
Tungsten lamp
assembly
Thumbscrew