Troubleshooting
Error Messages
48
Surveyor UV/Vis Plus Detector Hardware Manual
Thermo Scientific
Error Messages
Error messages are posted on the Status page of the data system that you are using to control
your detector.
Possible error messages include but are not limited to those listed in
.
Note
For further information, document the log entry and contact Thermo Fisher
Scientific Technical Support.
Table 8.
Possible error messages (Sheet 1 of 2)
Error number
Meaning
2001
System reset
2002
System error (bus/address)
2004
Lamp case open
2011
Sample light is too high
2012
Low light on D2 lamp
2013
Low light on W lamp
2014
Volatile RAM test failed
2015
Non-Volatile RAM test failed
2016
12 volts failed
2017
12 volt unregulated failed
2018
–12 volts failed
2019
Motor voltage failed
2020
W lamp voltage or driver failed
2021
D2 lamp or driver failed
2022
5 volts failed
2023
Reference diode failed
2024
Sample diode failed (Could be exposed to light.)
2025
Wavelength home test failed
2026
Second wavelength home failed
2027
Analog linearity test failed
2028
D2 lamp failed/not present
2029
W lamp failed/not present
2030
Shutter failed
2031
Unknown shutter/lamp failure