3
Maintenance
Lamp Maintenance
Thermo Scientific
Surveyor UV/Vis Plus Detector Hardware Manual
39
2. Open the front doors of the detector.
3. Using the 1/8 in. ball-end hex screwdriver (supplied), unscrew the captive retaining screw
on the lamp cover (see
).
Figure 28.
Lamp cover retaining screw
4. Pull the cover off to expose the lamp assemblies.
5. To read the chronometer, check the position of the electrolyte gap in the mercury tube
against the graduated background (see
).
Figure 29.
Chronometer for the deuterium lamp
CAUTION
Because the lamps run hot, turn off the power and allow sufficient time
(approximately 30 minutes) for the lamps to cool before you remove the lamp cover.
show the latest version of the
deuterium lamp.
Lamp cover retaining screw