Issue: 1.0
Operating Instructions
Thermo Fisher Scientific
SAM12 User Manual 5-53
- If a locked door is forced.
- If the back door is opened on a
two-door SAM12 following an
ALARM result.
In all cases, the Alarm will sound
at full volume until the door(s)
are closed.
Following this "ILLEGAL
DOOR SEQUENCE" message
and the doors have been shut, a
Residual Contamination Check
will be enforced before normal
operation can resume.
See
Contamination Check (page 5-
60)
for more information.
Press the START button to commence monitoring.
If the START button is not pressed within the Measurement
Start Timeout -see
for more information -
Normal Background Monitoring is resumed and SAM12
returns to the "Ready – Open door to Monitor Articles"
message.
See
Notes regarding Normal Background Monitoring (page 5-
for more information.
In this case, to re-start the monitoring of an article, the door
needs to be re-opened and shut again.
Once monitoring commences, the
SAM12 displays the "Counting"
message with an indication of the
measurement time remaining.
If the door is opened while the
measurement is in progress,
audible and visual warnings are
given and the measurement is
aborted. The "Measurement
Aborted – Please Close Front
Door" message is displayed.
Commence Monitoring
Содержание SAM12
Страница 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Страница 2: ......
Страница 4: ...This manual was produced using ComponentOne Doc To Help...
Страница 10: ......
Страница 18: ......
Страница 20: ......
Страница 34: ......
Страница 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Страница 130: ......
Страница 134: ......
Страница 144: ......
Страница 168: ......