Issue: 1.0
Operating Instructions
Thermo Fisher Scientific
SAM12 User Manual 5-61
If a door is opened during the
Residual Contamination Check,
the "ILLEGAL DOOR
SEQUENCE – Please close both
doors" message is displayed.
Once closed, the Residual
Contamination Count
recommences.
If the Instrument is contaminated,
it is not possible to open the
doors. Intervention by an
Administrator is necessary to
clear the contamination.
See
for more
information regarding clearing
the Contamination
Once the contamination has been cleared, and User Mode is re-
instated, the SAM12 performs the System Self Testing – see
.
Two-door SAM12
In a two-door SAM12, open the front door to remove a clean or
dirty Article – the "READY – DEPOSIT ARTICLE AND
CLOSE DOOR" message is displayed – see
(page 5-51)
. The SAM12 is now ready to monitor another
article.
On a two-door SAM12, if the
Rear Door is opened to remove a
clean article, the "REAR DOOR
IS OPEN – Please Wait" message
is displayed. Closing the door
displays the "Ready" message.
On a two-door SAM12, if the
Rear Door is opened to remove a
dirty article, the "ILLEGAL
DOOR SEQUENCE" message is
displayed. Closing the door
causes a Residual Contamination
Check to be run.
See
.
Instrument Contaminated
Remove Articles
Содержание SAM12
Страница 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Страница 2: ......
Страница 4: ...This manual was produced using ComponentOne Doc To Help...
Страница 10: ......
Страница 18: ......
Страница 20: ......
Страница 34: ......
Страница 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Страница 130: ......
Страница 134: ......
Страница 144: ......
Страница 168: ......