Issue: 1.0
Technical Description - Physics
Thermo Fisher Scientific
SAM12 User Manual 6-11
free space, (i.e. un-attenuated). The detectors were optimised
for each Isotope in the geometric centre of the measurement
cubicle.
Efficiencies are quoted “Centre” (geometric centre of
Measurement Volume).
and
3 inches (centre of Measurement Volume 3”
from floor)
Nominal Value quoted with maximum spread in brackets.
Door configuration has a small, usually insignificant, effect on
6 detector efficiencies.
Table 1 Standard SAM12 Standard Stainless Steel Liners
Variant
Total
Bgnd
(typ)
60
Co (1.2
MeV)
137
Cs (662
keV)
137
Ba (356
keV)
10
μ
R/h
0.1
μ
Sv/h
3 inches
Centre
3 inches
Centre
3 inches
Centre
4C-XD-
1L
350 cps
45%
(40-
48)
40%
(36-
44)
17%
(16-
20)
16%
(15-
19)
28%
(26-
31)
26%
(24-
28)
6C-2D-
1L
550 cps
57%
(55-
59)
53%
(51-
55)
24%
(22-
26)
23%
(21-
25)
38%
(36-
40)
36%
(34-
38)
6C-1D-
2L
350 cps
61%
(58-
65)
58%
(56-
60)
26%
(24-
28)
25%
(23-
27)
—
—
Low Energy Variants have 6 detectors, 2 inches (50 mm) lead
shielding and an additional 2mm Copper X-ray absorber.
(
Note
: mu-metal magnetic shielding is removed for maximum
sensitivity).
Table 2 Low Variant SAM12 Energy Variants
Total
Bgnd
241
Am (59 keV)
3 inches
Centre
SAM12LP – 6C-1D-2L
(Plastic Liners)
1000 cps
13½%
(12-15)
13%
(12-15)
SAM12LA – 6C-1D-2L
(Aluminium Liners)
1000 cps
9%
9%
SAM12A – Standard
Stainless Steel Liners
Low Energy Variants
Содержание SAM12
Страница 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Страница 2: ......
Страница 4: ...This manual was produced using ComponentOne Doc To Help...
Страница 10: ......
Страница 18: ......
Страница 20: ......
Страница 34: ......
Страница 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Страница 130: ......
Страница 134: ......
Страница 144: ......
Страница 168: ......