Operating Instructions
Issue: 1.0
5-62 SAM12 User Manual
Thermo Fisher Scientific
One-door SAM12
In a one-door SAM12, open the door to remove the Article –
the "READY – DEPOSIT ARTICLE AND CLOSE DOOR"
message is displayed - see
. The
SAM12 is now ready to monitor another article.
Out Of Service
If an error occurs during the
establishment of a new
background, the "Out of Service"
message is displayed (note that
the reason for the error may be
displayed in the message bar at
the bottom of the screen).
To clear this error, the state must
be cleared in the Administration
Mode (or will clear automatically
when the Out of Service
Recovery Interval is passed, see
for more
information).
See
for more
information regarding clearing
the Critical Error.
If High or Low Background occurs, the "Out of Service"
message is displayed.
Note that this Temporary Error message is recoverable.
The "Out Of Service –
Temporary Error has Occurred"
if High Background conditions
exist.
“Out of Service – High Background Conditions” message is
displayed when the background activity prevents discrimination
of the alarm level with the required confidence, in the
Maximum Monitoring Time allowed - see
. If the count rate from any detector exceeds 95% of the
Critical Error
Temporary Error
Notes regarding High
Background
Содержание SAM12
Страница 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Страница 2: ......
Страница 4: ...This manual was produced using ComponentOne Doc To Help...
Страница 10: ......
Страница 18: ......
Страница 20: ......
Страница 34: ......
Страница 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Страница 130: ......
Страница 134: ......
Страница 144: ......
Страница 168: ......