Issue: 1.0
Operating Instructions
Thermo Fisher Scientific
SAM12 User Manual 5-9
Select Set-Up|Params1 to
customise the first set of
Operational Parameters:
Description of options:
Field Name
Description
Low- Background
Alarm (cps)
If a background measurement falls below this value
(counts per second), a detector failure will occur.
Values: 0 to 99,999 cps (0 to disable)
Default: 50cps
Gross Changing
Background (
σ
)
This statistical variable is used by SAM12 to
determine if a changing background condition exists
on all detectors.
Values: 0.1 to 7 sigma in 0.1 sigma steps
Default: 5
σ
Detector Changing
Background (
σ
)
This statistical variable is used by SAM12 to
determine if a changing background condition exists
on a single detector.
Values: 0.1 to 7 sigma in 0.1 sigma steps
Default: 7
σ
Changing Conditions
(
σ
)
This statistical variable is used by SAM12 to
determine if a changing measurement condition
exists.
Values: 0.1 to 7 sigma in 0.1 sigma steps
Default: 7
σ
Changing Conditions
Period (s)
This is the length of time used by SAM12 to
determine if a changing measurement condition
exists.
Values: 2 to 30 s
Default: 3 s
RCC Contamination
Threshold (
σ
)
This is the statistical increase during a Residual
Contamination Check compared to the background
measurement.
Values: 0.1 to 7 sigma in 0.1 sigma steps
Default: 7
σ
Calibration Check
Accuracy (%)
This is the required accuracy to which the
calibration is carried out.
Default Calibration
Source
Use the pre-defined window to select a default
source. For a list of Sources, see
Cal Check (page
5-19)
.
Values: from source database
Calibration Required
Interval
This is the number of days since that last calibration
date before the unit needs to be calibrated again.
See
Out of Calibration (page 5-48)
to view the
Содержание SAM12
Страница 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Страница 2: ......
Страница 4: ...This manual was produced using ComponentOne Doc To Help...
Страница 10: ......
Страница 18: ......
Страница 20: ......
Страница 34: ......
Страница 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Страница 130: ......
Страница 134: ......
Страница 144: ......
Страница 168: ......