Operating Instructions
Issue: 1.0
5-28 SAM12 User Manual
Thermo Fisher Scientific
See
Performing the HV Scan (page 9-4)
for more information
regard the HV Scan.
Select Calibration|Thresholds to
set the Detector Thresholds:
This screen allows the individual voltage of any detector to be
viewed and updated.
To change any of the settings
, press in the field and update the
value. Press Apply Settings button to save or Cancel Changes
button to quit without saving.
Note that no threshold value should be set to a voltage that is
less than that of the preceding threshold. For example,
Threshold 5 must be set greater than or equal to Threshold 4.
Failure to follow this advice may invalidate measurements.
Please contact Thermo Fisher Scientific before changing
threshold values.
Select Calibration|Hv Power to
view the Amplifier Power
Settings
This screen allows the lower and upper limits for the Hv Power
to be viewed and updated. It measure the current being
delivered to the Hv Generator.
Select Calibration|Attenuation to
set a sequence to obtain
Attenuation Factors (note that a
Phantom object is required for
this process):
Thresholds
Hv Power
Attenuation
Содержание SAM12
Страница 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Страница 2: ......
Страница 4: ...This manual was produced using ComponentOne Doc To Help...
Страница 10: ......
Страница 18: ......
Страница 20: ......
Страница 34: ......
Страница 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Страница 130: ......
Страница 134: ......
Страница 144: ......
Страница 168: ......