Operating Instructions
Issue: 1.0
5-50 SAM12 User Manual
Thermo Fisher Scientific
When a new Background has
been established, the "Ready –
OPEN DOOR TO MONITOR
ARTICLES" message is
displayed. The SAM12 is now
ready to start monitoring articles.
If the front door is opened when
the SAM12 is establishing a new
background, background
counting stops, the "Close Door
To Resume" message is
displayed and an audible alarm
will sound. To continue with the
process, the door(s) must be shut.
If the back door is opened when
the SAM12 is establishing a new
background, background
counting stops, the "ILLEGAL
DOOR SEQUENCE" message is
displayed and an audible alarm
will sound. To continue with the
process, the door must be shut.
See
for more
information.
If the Residual Contamination
Check before Background Check
option is ticked (see
(page 5-6)
), a Residual
Contamination Count is
performed before a new Normal
Background Monitoring
commences.
See
Contamination Check (page 5-
60)
for more information.
If more than 15 minutes of
continuous operation have
elapsed since Normal
Background Monitoring has
commenced, a warning message
is displayed to advise the user
that a Background Check is
required.
In order for the Background
Check to be run before
Содержание SAM12
Страница 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Страница 2: ......
Страница 4: ...This manual was produced using ComponentOne Doc To Help...
Страница 10: ......
Страница 18: ......
Страница 20: ......
Страница 34: ......
Страница 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Страница 130: ......
Страница 134: ......
Страница 144: ......
Страница 168: ......