Operating Instructions
Issue: 1.0
5-60 SAM12 User Manual
Thermo Fisher Scientific
Contamination Check (page 5-
60)
for more information.
The Residual Contamination
Check is run under the following
circumstances:
- Following any door violation.
- If the Residual Contamination
Check before Background Check
option is set - see
.
- If Residual Contamination
Check after Alarm option is set -
see
.
The Residual Contamination
Check will verify that the current
background count rate does not
exceed the Background Count
Rate before the alarm, by a
statistically significant amount.
If no contamination is found, the
"Ready – OPEN DOOR TO
MONITOR ARTICLES"
message is displayed and the
SAM12 reverts to Background
mode.
However, if residual
contamination is detected or the
count rate from any detector
exceeds 95% of the Amplifier
“dead-time”, then an Instrument
Contaminated state exists. The
"INSTRUMENT
CONTAMINATED – DO NOT
USE" message is displayed (see
Instrument Contaminated (page
5-61)
for more information).
Note that if the Door Locked
options are ticked, the door(s) are
locked while the Instrument is
Contaminated.
See
for more
information regarding Lock Front
Door on Alarm.
Notes regarding Residual
Contamination Check
Содержание SAM12
Страница 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Страница 2: ......
Страница 4: ...This manual was produced using ComponentOne Doc To Help...
Страница 10: ......
Страница 18: ......
Страница 20: ......
Страница 34: ......
Страница 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Страница 130: ......
Страница 134: ......
Страница 144: ......
Страница 168: ......