Chapter 7
|
Data Transfer and Software Updates
Thermo Scientific
Orion Star A210 Series Benchtop Meter User Manual
|
93
10.
Click the Upgrade Firmware button. Wait while the update is installed on the meter.
a.
Warning: The software update must not be interrupted once the Update Firmware
button is clicked. Interrupting the software update may corrupt the meter and cause it to
not power on correctly.
11.
The update will take approximately 15 minutes. Once the software update is complete, the
dialog box will show “Done!”
12.
Close the software update and detach the cable from the meter and computer.
Note: If the following message is shown and the USB
cable is being used, turn the power to the meter off by
unplugging the power adapter and/or remove batteries,
disconnect the USB cable, close the software update
and retry this procedure using the RS232 cable.
Содержание Orion Star A210 Series
Страница 5: ......