Chapter 3
|
Meter Setup Menus
Instrument Settings Setup Menu
Use the Instrument Settings menu to update meter settings for data transfer, data log, date and
time, language, sounds, stirrer speed, display contrast, auto-shutoff, user ID and sample ID.
Instrument
Setting
Options
Details
Export Data
•
Printing
Off
On
•
Comm Setup
RS232 USB
•
Data Format
Printer PC (CSV)
•
Comm Config
1200
9600
2400
19200
4800
38400
Access the data export menu to set data transfer
settings for exporting measurement and calibration
information to a printer or computer
Turn printing on to allow data transfer from the meter to
an external device, select the data transfer format as
printer (standard text) or PC (comma delimited text)
and set the meter baud rate to match the printer or
computer baud rate
The meter will automatically detect when an RS232 or
USB cable is connected to a meter port
Data Log
•
Off
•
On
Turn data logging on to save measurement data to the
data log according to the selected read type
Date and Time
•
Date
DD/MM/YY
MM/DD/YY
Date: 01/01/12
•
Time
12 Hour Clock
Time: 02:30 PM
24 Hour Clock
Time: 14:30
Use the date setting to select the date format as
day/month/year (DD/MM/YY) or month/day/year
(MM/DD/YY) and then enter the month, day and year
values
Use the time setting to set the time format as 12 hour
clock or 24 hour clock and then enter the hour and
minute values
Language
•
English
•
Español
•
Deutsch
•
Français
•
Italiano
•
中文
(Chinese)
Set the language to be used for the meter interface,
additional languages (including Portuguese and
Korean) are available with software updates on our
website at
www.thermoscientific.com/OrionMeters
Key Press Beep
•
Off
•
On
Turn the audible beep on or off each time a meter key
is pressed
Alarm Beep
•
Off
•
On
Turn the audible beep on or off when an alarm is
triggered
Stirrer
Adjust the stirrer speed setting from one (slowest) to
five (fastest) when using an Orion Star stirrer probe
Contrast
Adjust the display contrast higher or lower to improve
display viewing in different light conditions
Auto Shut Off
•
Off
•
On
Turn on function to automatically power off the meter
when no keys are pressed for 20 minutes
User ID
ABCDE
Enter a user ID value (up to six characters) using the
alphanumeric entry screen
Sample ID
•
Off
•
Manual
•
Auto Increment
Set sample ID as off, manual (up to six alphanumeric
characters) or auto incremental (up to six numbers that
are automatically increased for each measurement)