Chapter 8
|
Customer Services
Orion Star A213 RDO/DO Benchtop Meter Specifications
Measurement Channels
1 – dissolved oxygen as % saturation or mg/L with temperature
Dissolved
Oxygen
Polarographic
Concentration
% Saturation
Range
0 to 90 mg/L
0 to 600 %
Resolution
0.01 / 0.1 mg/L
0.1 / 1 %
Relative Accuracy ±0.2 mg/L or ±2% of reading,
whichever is greater
±2 % saturation or ±2% of
reading, whichever is greater
RDO Optical
Concentration
% Saturation
Range
0 to 50 mg/L
0 to 500 %
Resolution
0.01 / 0.1 mg/L
0.1 / 1 %
Relative Accuracy
±0.1 mg/L up to 8 mg/L,
±0.2 mg/L 8 to 20 mg/L,
±10% of reading up to 50 mg/L
±2 % saturation ≤ 200 %
saturation, ±10 % saturation
> 200 % saturation
Barometric Pressure Correction
400 to 850 mmHg, automatic using built-in barometer (±6 mmHg)
or manual entry
Salinity Correction Factor
0.0 to 45.0 ppt, automatic using manual entry of sample salinity
Calibration Types
Water-saturated air, air-saturated water, manual (Winkler), zero
point
Compatible Sensors
Polarographic, RDO optical
Temperature
Range
0 to 50 °C, 32 to 122 °F
Resolution
0.1 °C, 0.1 °F
Relative Accuracy ±0.1 °C
Offset Calibration
1 point
Source Options
Automatic with built-in temperature sensor
Sensor Input 9 pin MiniDIN
Dissolved oxygen sensor with built-in temperature
102
|
Orion Star A210 Series Benchtop Meter User Manual
Thermo Scientific