Chapter 7
|
Data Transfer and Software Updates
Cal Mode
PC (CSV) Format
DO
(Air or Water
Calibration )
Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &
Time, Calibration Point, Calibration Type, Calibration Saturation Value, Calibration
Saturation Unit, Set Zero Calibration Current Value, Set Zero Calibration Current Unit,
Calibration Current Value, Calibration Current Unit, Solution Temperature Value,
Solution Temperature Unit, Membrane Temperature Value, Membrane Temperature
Unit, Calibration Pressure Value, Calibration Pressure Unit, Salinity Value, Salinity Unit,
Calibration Slope Value, Calibration Slope Unit, Calibration Point, Calibration Number
Example:
A213 DO/RDO, X01039, 3.04, CH-1, DO, 01/03/15 16:05:41, 1, Auto- Air, 102.3,%, 0.0,
nA, 405.1, nA, 5.0, C, 5.0, C, 736.1, mmHg, 0.0, ppt, 4.1, nA/%sat, 1, #1 <CR>
DO
(Manual
Calibration )
Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &
Time, Calibration Point, Calibration Type, Calibration Concentration Value, Calibration
Concentration Unit, Set Zero Calibration Current Value, Set Zero Calibration Current
Unit, Calibration Current Value, Calibration Current Unit, Solution Temperature Value,
Solution Temperature Unit, Membrane Temperature Value, Membrane Temperature
Unit, Calibration Pressure Value, Calibration Pressure Unit, Salinity Value, Salinity Unit,
Calibration Slope Value, Calibration Slope Unit, Calibration Point, Calibration Number
Example:
A213 DO/RDO, X01039, 3.04, CH-1, DO, 01/03/15 16:05:41, 1, Manual, 12.62, mg/l, 0.0,
nA, 405.1, nA, 5.0, C, 5.0, C, 736.1, mmHg, 0.0, ppt, 4.1, nA/%sat, 1, #2 <CR>
RDO
(Air or Water
Calibration)
Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &
Time, Probe Serial Number, Calibration Point, Calibration Type, Calibration Saturation
Value, Calibration Saturation Unit, Set Zero Partial Pressure Value, Set Zero Partial
Pressure Unit, Calibration Partial Pressure Value, Calibration Partial Pressure Unit,
Solution Temperature Value, Solution Temperature Unit, Barometric Pressure Value,
Barometric Pressure Unit, Salinity Value, Salinity Unit, Calibration Slope Value,
Calibration Slope Unit, Calibration Point, Calibration Number
Example:
A213 DO/RDO, X01039, 3.04, CH-1, RDO, 01/03/15 16:05:41, 123454, 1, Auto–Air,
100.0,%, 8.2, Torr, 149.6, Torr, 24.3, C(ATC), 749.7, mmHg, 0.1, ppt, 1.52, Torr/%sat, 1,
#1 <CR>
RDO
(Manual
Calibration)
Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &
Time, Probe Serial Number, Calibration Point, Calibration Type, Calibration
Concentration Value, Calibration Concentration Unit, Set Zero Partial Pressure Value,
Set Zero Partial Pressure Unit, Calibration Partial Pressure Value, Calibration Partial
Pressure Unit, Solution Temperature Value, Solution Temperature Unit, Barometric
Pressure Value, Barometric Pressure Unit, Salinity Value, Salinity Unit, Calibration Slope
Value, Calibration Slope Unit, Calibration Point, Calibration Number
Example:
A213 DO/RDO, X01039, 3.04, CH-1, RDO, 12/16/12, 09:21:00, 123454, 1, Manual, 6.9,
mg/l, 8.2, Torr, 149.6, Torr, 24.3, C(ATC), 749.7, mmHg, 0.1, ppt, 1.52, Torr/%sat, 1, #1
<CR>
Thermo Scientific
Orion Star A210 Series Benchtop Meter User Manual
|
89