Chapter 3
|
Meter Setup Menus
Cell K Setting
Orion Star A212 conductivity meters and Orion Star A215 pH/conductivity meters are capable
of automatically recognizing Thermo Scientific Orion 100 μS conductivity standard, 1413 μS
conductivity standard and 12.9 mS conductivity standard when the nominal cell constant (K)
value of the conductivity sensor is entered in the setup menu.
Catalog Number
Description
011008
Orion 100 μS conductivity standard, 5 x 60 mL
011007
Orion 1413 μS conductivity standard, 5 x 60 mL
01100710
Orion 1413 μS conductivity standard, 10 individual use pouches
011006
Orion 12.9 mS conductivity standard, 5 x 60 mL
01100610
Orion 12.9 mS conductivity standard, 10 individual use pouches
The nominal cell constant (K) values for Thermo Scientific Orion conductivity sensors
compatible with Star A210 series conductivity meters are listed below.
Catalog
Number
Description
Measurement Range
Nominal Cell
Constant
013005MD Orion 4-cell conductivity/temperature sensor with
1.5 meter (5 foot) cable
1 μS/cm to 200 mS/cm
0.475 cm
-1
013010MD Orion 4-cell conductivity/temperature sensor with
3 meter (10 foot) cable
1 μS/cm to 200 mS/cm
0.475 cm
-1
013020MD Orion 4-cell conductivity/temperature sensor with
6 meter (20 foot) cable
1 μS/cm to 200 mS/cm
0.475 cm
-1
013025MD Orion 4-cell conductivity/temperature sensor with
10 meter (32 foot) cable
1 μS/cm to 200 mS/cm
0.475 cm
-1
013605MD Orion 4-cell conductivity/temperature sensor with
1.5 (5 foot)meter cable
10 μS/cm to 200 mS/cm 0.55 cm
-1
013610MD Orion 4-cell conductivity/temperature sensor with
3 meter (10 foot) cable
10 μS/cm to 200 mS/cm 0.55 cm
-1
013016MD Orion pure water 2-cell conductivity/temperature
sensor with 1.5 meter (5 foot) cable
0.01 μS/cm to 300 μS/cm 0.1 cm
-1
011510MD Orion 2-cell conductivity/temperature sensor with
3 meter (10 foot) cable
10 μS/cm to 200 mS/cm 1.0 cm
-1
011050MD Orion 2-cell conductivity/temperature sensor with
1.5 meter (5 foot) cable
1 μS/cm to 20 mS/cm
1.0 cm
-1
018020MD Orion high range 2-cell conductivity sensor with
1.5 meter (5 foot) cable
10 μS/cm to 2000 mS/cm 10 cm
-1
36
|
Orion Star A210 Series Benchtop Meter User Manual
Thermo Scientific
Содержание Orion Star A210 Series
Страница 5: ......