Chapter 2
|
Meter Basics
Attaching the Electrode Stand and Holder
The electrode stand can be attached to either side of the meter and up to two stands can be
attached to each meter. A weighted base (Catalog Number STARA-HB) is also available to
support the stand without attachment to the meter.
1.
Open the box containing the electrode stand. The box will include a base plate, electrode
arm, pivot pin, electrode holder, ATC spacer and RDO clip.
2.
Turn the meter over, with the meter display facing down, on a clean dry surface.
3.
Identify the side of the meter that the stand will be installed on and remove the screw
between the circles on that side of the meter.
4.
Align the base plate of the stand with the circles on the meter.
5.
Replace the screw from step 3 to attach the base plate to the meter.
6.
Turn the meter over, with the meter display facing up.
7.
Insert the electrode arm into the metal post on the base plate.
8.
Connect the electrode holder to the electrode arm using the pivot pin.
10
|
Orion Star A210 Series Benchtop Meter User Manual
Thermo Scientific