Chapter 7
|
Data Transfer and Software Updates
Meter Calibration Data from GETCAL Command
Cal Mode
PC (CSV) Format
pH
Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &
Time, Total Calibration Points,
repeated for each calibration point: Calibration Point, pH Value, pH Unit, mV Value,
mV Unit, Temperature Value, Temperature Unit,
repeated for each point-to-point segment: Slope Value, Slope Unit,
repeated for each point-to-point segment: Offset Value, Offset Unit,
Average Slope Value, Average Slope Unit, Calibration Number
Example (5 point calibration):
A211 pH, X01036, 3.04, CH-1, pH, 01/03/15 16:05:41, 5, 1, 2.00, pH, 290.5, mV, 25.0,
C, 2, 4.01, pH, 174.3, mV, 25.0, C, 3, 7.00, pH, 0.0, mV, 25.0, C, 4, 10.01, pH, -173.9,
mV, 25.0, C, 5, 12.00, pH, -289.7, mV, 25.0, C, 97.7,%, 98.5,%, 97.7,%, 98.4,%, 1.4,
mV, 0.0, mV, 0.0, mV, 1.3, mV, 98.1,%, #1 <CR>
RmV
Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &
Time, Raw mV Value, Raw mV Unit, Relative mV Offset Value, Relative mV Offset Unit,
Reference mV Value, Reference mV Unit, Calibration Number
Example:
A211 pH, X01036, 3.04, CH-1, RmV, 01/03/15 16:05:41, 0.0, mV, 0.0, mV, 0.0, mV, #1
<CR>
ORP
Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &
Time, Raw mV Value, Raw mV Unit, Offset Value, Offset Unit, Temperature Value,
Temperature Unit, Calibration Number
Example:
A211 pH, X01036, 3.04, CH-1, ORP, 01/03/15 16:05:41, 176.9, mV, 242.6, mV, 25.0, C,
#1 <CR>
ISE
Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &
Time, Total Calibration Points,
repeated for each calibration point: Calibration Point, Concentration Value,
Concentration Unit, mV Value, mV Unit, Temperature Value, Temperature Unit,
repeated for each point-to-point segment: Slope Value, Slope Unit,
repeated for each point-to-point segment: Offset Value, Offset Unit,
Average Slope Value, Average Slope Unit, Blank Value, Blank Unit, Calibration Number
Example (5 point calibration):
A214 pH/ISE, X01036, 3.04, CH-1, ISE, 01/03/15 16:05:41, 5, 1, -1, ppb, 0.0, mV, 25.0,
C, 2, 2, ppb, 19.0, mV, 25.0, C, 3, 300, ppb, 150.1, mV, 25.0, C, 4, -2000, ppb, 200.1,
mV, 25.0, C, 5, 4000, ppb, 210.1, mV, 25.0, C, 63.1, mV/dec, 60.2, mV/dec, 60.7,
mV/dec, 33.2, mV/dec, 0.0, mV, 0.9mV, 0.0mV, -0.2mV, 90.4, mV, 54.3, mV/dec,
0.0000, ppb, #1 <CR>
Conductivity
Meter Model, Serial Number, Software Revision, Channel, Mode, Calibration Date &
Time, Total Calibration Points,
repeated for each calibration point: Calibration Point, Conductivity Value,
Conductivity Unit, Conductance Value, Conductance Unit, Temperature Value,
Temperature Unit (MTC/ATC), Calibration Type, Calibration Factor,
Average Cell Constant (K), Calibration Number
Example (5 point calibration):
A212 Cond, X01038, 3.04, CH-1, Cond, 01/03/15 16:05:41, 5, 1, 52.8, uS/cm, 111.1, uS,
25.0, C, Manual, 0.4750, 2, 68.6, uS/cm, 143.1, uS, 25.0, C, Manual, 0.4956, 3, 100,
uS/cm, 200.2, uS, 25.0, C, Manual, 0.5511, 4, 158.4, uS/cm, 333.4, uS, 25.0, C, Manual,
0.4375, 5, 475, uS/cm, 1000, uS, 25.0, C, Manual, 0.4749, 0.4868 #1 <CR>