Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Chapter V
Instrument Maintenance
The .Thermo .Scientific .Orion .2111LL .sodium .monitor .is .designed .for .
simple .maintenance . . .Follow .the .instructions .in .this .section .to .ensure .
proper .operation .of .your .instrument .
To .ensure .proper .maintenance .and .good .monitor .performance, .a .service .
logbook .should .be .kept .
• . Record .maintenance .date .and .type .service .work .completed . .
• . Mark .date .and .fluid .levels .of .reagent .and .date .when .replaced . .
• . Tag .each .electrode .cable .with .installation .date .of .electrode .
1 . . . Check .that .sample .flow .rate .is .between .35 .and .45 .mL/min . . .To .alter .
flow .rate, .pull .out .locking .knob .of .pressure .regulator, .then .rotate .black .
knob .to .increase .(clockwise) .or .decrease .(counterclockwise) .sample .
flow . . .Push .in .to .lock .rate, .if .desired . .
2 . . . Inspect .unit .for .leakage . . .Diffusion .tubing .leakage .is .indicated .by .a .rise .
in .level .of .reagent . .
3 . . . Check .that .there .are .no .error .indications .and .current .concentration .
level .is .reasonable . .
4 . . . Check .that .reference .fill .solution .is .adequate . .
5 . . . If .required .wipe .instrument .with .damp .cloth .to .remove .dirt .and . .
dust .particles . .
Maintenance Schedule
Weekly Maintenance
Recommendation