II-
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
4 . . Replace .large .single .O-ring .in .reagent .bottle .adapter .assembly . . .Use .
needle .nose .pliers .if .necessary .
5 . . Replace .all .3 .O-rings .on .face .of .reagent .manifold . . .Refer .to . .
Figure II-11
.
6 . . Slide .the .reagent .bottle .adapter .assembly .over .the .threaded .screw .of .the .
reagent .manifold . . .Refer .to .
Figure II-11
.
7 . . Tighten .the .thumbnut .to .secure .
Figure II-
Reagent Manifold & Reagent
Bottle Adapter Assembly
(3) O-rings for Manifold
Manifold Screw
O-ring for Bottle Adapter
Diffusion Tubing
Reagent Bottle
Reagent Bottle Adapter
Reagent Manifold