V-6
Instrument Maintenance
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Using .the .restrictor .tube .assembly .(Cat . .No . .2111RT) .from .the .Orion .
annual .maintenance .kit .(Cat . .No . .2111MK) . .Refer .to .
Figure V-4
.for .
restrictor .tubing .connections .
1 . . Close .sample .inlet .valve .to .stop .sample .flow .through .monitor .by .
turning .valve .to .the .right .
2 . . Loosen .restrictor .tube .fitting .connection .from .top .of .calibration .shut .
off .valve .
3 . . .Loosen .restrictor .tube .fitting .from .reagent .manifold .
4 . . Pull .out .the .restrictor .tube .assembly .at .each .end . .
5 . . Unscrew .clip-holding .restrictor .to .panel .
6 . . Replace .with .new .restrictor .tubing .assembly . .
7 . . Be .sure .all .fittings .are .secure, .to .prevent .leaks .
8 . . Open .sample .inlet .valve .to .start .sample .flow .through .monitor .by .
turning .valve .to .the .left .
Replacement of Restrictor
Tubing
Figure V-4
Restrictor Tubing
50
40
30
20
10
Reagent Bottle
Air Pump
Right Air Valve
Restrictor Tube
Assembly