II-9
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Warning:
Be .sure .to .read .and .observe .the .following .requirements! . .
Diisopropylamine .is .hazardous . . .Refer .to .MSDS .for .DIPA, .available .at .
www .thermo .com/water . . .Use .protective .glasses .and .gloves . . .Refer .to .bottle .
label .for .precautions . . .Installation .of .DIPA .reagent .requires .a .fume .hood .or .
well-ventilated .area . . .
s
If .monitor .is .offline .or .sample .is .shut .off .from .the .instrument .for .a .
prolonged .period .of .time .be .sure .to .follow .the .
Chapter III, Shutdown &
Start-Up
.
Procedure
.
1 . . Unscrew .thumbnut .and .remove .reagent .bottle .from .the .reagent .
manifold . .
• . .New .instruments .will .ship .with .a .rubber .coated .glass .reagent .bottle .
and .complete .diffusion .tubing .assembly .installed . . .Only .installation .
of .reagent .is .necessary .for .first .time .start .up .
2 . . Twist .open .reagent .bottle, .separating .glass .bottle, .reagent .bottle .adapter .
assembly .and .plastic .cap .to .reagent .bottle .
3 . . Unscrew .the .fittings .from .base .of .the .reagent .bottle .adapter .assembly . .
dispose .of .used .diffusion .tubing .assembly .accordingly .
The .new .diffusion .tubing .assembly .for .use .with .the .Orion .2111LL .low .
level .sodium .monitor .provides .superior .locking .ability .to .maximize .uptime .
and .prevent .against .costly .blowouts . . .The .newly .designed .HDPE .locking .
mechanism .is .precisely .engineered .to .remain .connected .under .variable .
temperature .conditions .and .pressure .fluctuations . . .The .new .style .diffusion .
tubing .assembly .is .now .suspended .in .the .headspace .of .the .reagent .bottle- .
decreasing .the .rate .of .consumption .of .reagent-saving .you .time .and .money .
The .diffusion .tubing .ships .fully .assembled .for .quick .and .easy .installation . .
Occasionally .the .assembly .will .loosen .during .shipment . .
• . Be .sure .to .check .the .diffusion .assembly .for .completeness .of .
components .prior .to .installation .upon .the .reagent .bottle .cap . .
• . Do .not .use .Teflon .tape .during .installation
• . Twisting .of .the .clear .silicone .tubing .will .compromise .tubing .integrity .
and .may .cause .ruptures . . .Be .careful .to .avoid .kinks .and .twisting .during .
installation .
Installation of Reagent
& Diffusion Tubing
Recommendation
Diffusion Tubing Assembly
Tips Before Installation
Part Description
Quantity
O-ring .seal .fitting
2
Twist .and .lock .Cap
2
Pipe .Fitting
2
O-ring .EPDM
6
Ferrule
4
Silicone .Tubing
1