IV-
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage
User Action
Marquee
Main Display
top line
bottom line
Valve position
3
Press
.
3
System will proceed to measurement
mode
Actual temperature
reading
Actual
readings in
ppb & mV
System enters measurement
mode, but will remain on hold
until hold timer expires.
(Default 30 minutes)
Calibration Abort- Exit Steps
Exit
ARE YOU SURE?
PRESS ENTER TO
ABORT OR PRESS
LAST SCREEN TO
BACKUP.
SUrE?
Abort calibration vs. Last
Screen option:
The system allows the user
two options when
is
pressed at any step during the
calibration.
Go to
Exit
Exit
Decide whether or not you really want to
exit the calibration process and lose the
new slope and E
0
:
To
Abort
calibration:
Follow step
Exit 3
:
User enters into the measure mode.
User must refer to
stage 30
for proper air
and fluid flow for measurement mode.
To
Continue
calibration:
If
is pressed, system will go back to
last screen used in calibration sequence
prior to pressing
.
Continue to follow calibration according to
Marquee and main display directions.
Exit 3
Press
.
Exit 4
CALIBRATION
ABORTED
PRESS ENTER
CAL Abrt
Exit 5
Press
.
Refer to
stage 30
for proper air
and fluid flow for measurement
mode. System return will return
to measure mode.
(Begin again with
stage
to
retry calibration sequence)