IV-5
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
The .offline .calibration .feature .is .simply .a .one-point .calibration .without .
slope .change . .The .term .“offline .calibration” .refers .only .to .the .fact .that .
a .sample .from .2111LL .bypass .is .taken .“offline” .for .laboratory .analysis; .
in .fact, .no .downtime .is .experienced .and .the .instrument .remains .online .
throughout .the .procedure .
Offline Calibration
Procedure
Stage
User Action
Marquee
Main Display
top line
bottom line
Notes
Open bypass valve to redirect
flow.
Take QC sample from bypass
of 2111LL.
Actual temperature
reading
Actual readings in
concentration &
mV
*
Press
to start calibration
*
If calibration password is
activated - user must scroll
in digits 0-9 using
.
./ .
.
to set digits and
to move selected digit.
Press
to accept
password.
ENTER PASSWORD PASS
0000
CAL appears as Mode.
3
DKA or OFFL
(Flashing)
CAL
tYPE
4
Press
.
./ .
.
arrows
to select OFFL
OFFL
(Flashing)
CAL
tYPE
5
Press
.
HOLD CAL appears as mode.
6
Wait 0 - 5 minutes
AWAITING DATA
STABILITY
Actual readings in
ppb & mV flashing
System continuously evaluates data by
applying stability criteria until a “stable/
pass” condition is met.
When stability is reached the values
are stored and the system automatically
returns to the measurement mode.
7
Perform laboratory analysis
on QC sample and return with
results
System remains in measurement mode
and operates as normal. The CAL and
MEASURE mode icons will both be
displayed.