IV-8
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage
User Action
Marquee
Main Display
top line
bottom line
Valve position / Notes
7
Press
.
8
Wait 7 minutes
STARTING CALIBRATION
MEASURING BACKGROUND
WAIT 7 MINUTES
Actual readings in
ppb & mV flashing
System circulates sample.
Minutes will be counted down
on Marquee.
9
Wait 0 - 5 minutes
AWAITING DATA STABILITY
Actual readings in
ppb & mV flashing
System continuously evaluates
data by applying stability criteria
until a “stable/pass” condition is
met. If system stabilizes follow
stage 0
.
If system stability is not reached
a timeout occurs. Error E041 is
issued. Jump to step
Error E04
.
0
Add Std 1 through cal port
ADD STD1
PRESS ENTER
Add Std1
Press
.
Wait 5 minutes
MEASURING STD1
CIRCULATING
WAIT 5 MINUTES
Actual readings in
ppb & mV flashing
System circulates sample.
Minutes will be counted down
on Marquee.
3
Wait 0 - 5 minutes
AWAITING DATA STABILITY
Actual readings in
ppb & mV flashing
System continuously evaluates
data by applying stability criteria
until a “stable/pass” condition is
met. If system stabilizes follow
stage 4
.
If system stability is not reached
a timeout occurs. Error E041 is
issued. Jump to step
Error E04
.
4
Add Std 2 through cal port
ADD STD2
PRESS ENTER
Add Std2
5
Press
.
6
Wait 4 minutes
MEASURING STD2
CIRCULATING
WAIT 4 MINUTES
Actual readings in
ppb & mV flashing
System circulates sample.
Minutes will be counted down
on Marquee.