IV-6
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage
User Action
Marquee
Main Display
top line
bottom line
Notes
8*
Press
.
.
* If calibration password is
activated - user must scroll
in digits 0-9 using
.
./ .
.
to set digits and
to move selected digit.
Press
to accept
password.
ENTER PASSWORD PASS
0000
CAL appears as Mode
9
DKA or OFFL
(Flashing)
CAL
tYPE
0
Press
.
./ .
to select
OFFL.
OFFL
(Flashing)
CAL
tYPE
Press
.
HOLD CAL appears as mode
OFFLINE CAL
CONC
# #.#
(Flashing)
3
Adjust the displayed value to
the QC value obtained from
laboratory analysis.
Use
to set decimal and
ppm / ppb.
Use
.
./ .
to set the
first digit.
Use
to move to set each
digit.
Use
.
./ .
to set the
values of remaining digits.
OFFLINE CAL
CONC
#
#.#
(Flashing digit will
change)
4
Press
.
The system recalculates the E
0
and returns
to the measurement mode.
Only the MEASURE mode icon should be
displayed.