System Installation, Set Up and Shut Down
Recommended Initial Parameter Settings
Thermo Fisher Scientific
Mercury System Manual
2-7
Table 2–4.
80
i
Alarms Settings
Parameter
Range
Typical Value
Internal Temperature
15 – 45 °C
32 °C
Chamber Temperature
40 – 50 °C
45 °C
Cooler Temperature
10 – 15 °C
14 °C
Pressure
25 – 75 mmHg
50 mmHg
Flow
0.000 – 1.000 LPM
0.30 LPM
Intensity
10,000 – 150,000 Hz 80,000 Hz
Zero Check (Analyzer Max Offset)
0 – 600 µg
2.0 µg
Span Check (Analyzer Max Offset) 0 – 600 µg
4.0 µg
Zero AutoCal (Probe Max Offset)
0 – 600 µg
2.0 µg
Span AutoCal (Probe Max Offset)
0 – 600 µg
4.0 µg
Probe Dilution Factor
1 – 100
30
Hg(0) Concentration
0 – 600 µg
600 µg
Hg(2+) Concentration
0 – 600 µg
600 µg
Hg(t) Concentration
0 – 600 µg
600 µg
Mother Board Status
OK
Interface Board Status
OK
Table 2–5.
Probe Alarms Settings
Parameter
Range
Typical Value
Umbilical Temperature
100 – 200 °C
120 °C
Probe Temperature
190 – 250 °C
220 °C
Converter Temperature 700 – 900 °C
760 °C
Oxidizer Temperature
360 – 440 °C
400 °C
Orifice Pressure
0.0 – 3.0 psig
0.5 psig
Dilution Pressure
0.0 – 70.0 psig
50 psig 83
i
, 35 psig 83
i
GC
Blowback Pressure
0.0 – 70.0 psig
60 psig
Eductor Pressure
0.0 – 30.0 psig
10 psig
Vacuum Pressure
0.0 – 29.0 inHg 21.0 inHg 83i, 18 inHg 83
i
GC
80
i
Alarms Settings
Probe Alarms Settings
Содержание Mercury Freedom 80i
Страница 1: ...Mercury System Manual Part Number 105933 00 28Jan2008 ...
Страница 6: ......
Страница 10: ......
Страница 12: ......
Страница 14: ......
Страница 42: ......
Страница 66: ......
Страница 79: ...Optional Equipment Nitrogen Generator Thermo Fisher Scientific Mercury System Manual 6 9 ...
Страница 100: ......