Introduction
Model 83i
Thermo Fisher Scientific
Mercury System Manual
1-7
regulators, as well as an electronic vacuum transducer monitoring the
dilution module vacuum in the Model 83
i
.
The Model 83
i
extraction probe includes an filter with a built in dilution
module and high temperature thermal converter for reducing oxidized Hg
to elemental for subsequent analysis by the Model 80
i
.
The Model 82
i
provides both 220 and 110 AC voltage to the Model 83
i
.
220 VAC powers the probe’s stinger heater and filter/eductor heater. 110
VAC powers the Total Hg converter as well as four (4) probe solenoid
valves for cal/zero gas, stinger blowback, filter blowback, and Hg spike. The
Model 82
i
also receives two (2) 4-20 ma signals from the probe critical
orifice pressure transducer and the fast loop venturi pressure transducer.