System Installation, Set Up and Shut Down
Recommended Initial Parameter Settings
2-6
Mercury System Manual
Thermo Fisher Scientific
This section provides recommendations for initial parameter settings.
Table 2–2
lists the initial service settings.
Table 2–2.
Initial Service Settings
Parameter
Typical Range
PMT Voltage
500 – 1000 V
Lamp Intensity
50k – 150k Hz
Cooler Temperature
10 – 15 °C
Bench Temperature
45 – 50 °C
Probe Temperature
190 – 220 °C
Valve Failsafe Temperature 190 °C
Umbilical Temperature
120 °C
Converter Temperature
760 °C
Blowback Pressure
40 – 65 psig
Eductor Pressure
5 – 20 psig
Dilution Pressure
30 – 65 psig
Oxidizer Temperature
400 °C
Table 2–3.
Set-up Blowback Schedule
Parameter
Range
Default
Period
0 - 24 Hours
12 Hours
Filter Duration
0 - 60 Seconds
10 Seconds
Stinger Duration 0 – 60 Seconds 10 Seconds
Recommended Initial
Parameter Settings
Initial Service Settings
Set-up Blowback
Schedule