System Installation, Set Up and Shut Down
Plumbing and Electrical Hookup
2-2
Mercury System Manual
Thermo Fisher Scientific
Refer to the “System Reference Drawings” in Appendix B for the system
plumbing and electrical hookup connections.
Equipment Damage
To prevent damaging the probe transducer, read the
cable labels carefully and DO NOT connect the C or D cable to the wrong
connector on the 82
i
rear panel.
▲
Refer to
Table 2–1
for setting the default temperatures (or per site
requirements). At the 80
i
Main Menu, select > Service >
Set
Temperatures
:
Table 2–1.
Default Temperatures
Parameter
Default Temperature
Setting
Bench Temp
45 °C
Probe Temp
220 °C
Umbilical Temp
120 °C
Converter Temp
760 °C
Failsafe Valve Temp
200 °C
Oxidizer Temp
400 °C
Use the following procedure to power up the Mercury CEMS.
Note
If the 81
i
is in Service Mode, it will ignore 80
i
commands.
▲
1.
Apply Power to the CEMS, turn power switches ON in the following
order:
Model 82
i
Model 81
i
2.
After the Model 81
i
initializes and the Run screen is displayed, power
the Model 80
i
.
3.
To test communication from the 80
i
to the 81
i
:
a.
Using the Model 80
i
, change the Gas Mode to one of the
following: Instrument Zero, Instrument Cal, System Zero, or
System Cal.
Plumbing and
Electrical Hookup
Set Temperature
Defaults
Mercury CEMS
Power-Up Sequence
Содержание Mercury Freedom 80i
Страница 1: ...Mercury System Manual Part Number 105933 00 28Jan2008 ...
Страница 6: ......
Страница 10: ......
Страница 12: ......
Страница 14: ......
Страница 42: ......
Страница 66: ......
Страница 79: ...Optional Equipment Nitrogen Generator Thermo Fisher Scientific Mercury System Manual 6 9 ...
Страница 100: ......