Optional Equipment
Hydrator
Thermo Fisher Scientific
Mercury System Manual
6-5
1.
Remove 3/8-inch Teflon Umbilical Line 3, (if previously installed),
from “PROBE” output of 81
i
. Attach this line to the output of the
Hydrator. This is the 3/8-inch Kynar fitting on the right of the
Hydrator when viewed from the front.
Note
If performing a field retrofit and the umbilical line was cut short, use
the 3/8-inch Teflon tubing and Kynar union fittings supplied with the
retrofit kit to reach the Hydrator output fitting.
▲
2.
Take enough 3/8-inch Teflon tubing to run from the input (left side as
viewed from front) of the Hydrator to the 3/8 – 1/4-inch reducing
union where Umbilical Line 3 was prior to relocation during Step 1.
3.
Open the 81
i
and flip the check valve 180 degrees so that the arrow is
facing the back of the instrument.
4.
Check tightness of the fittings internal and external to the Hydrator,
and apply air to system by opening the air handling inlet valve.
1.
At the 80
i
Main Menu, select Instrument Controls > I/O
Configuration >
Output Relay Settings
, and press
.
2.
With the arrow keys, select #6, and press
.
3.
Press
again to set the Logic State to “NCL.”
4.
Select Instrument State, Choose Signal Type,
Non-Alarm
, then select
Sample Mode located at the
bottom
of the list.
5.
Return to the I/O Configuration menu, Digital Input Settings and
press
.
6.
With the arrow keys select #7, and press
.
7.
Select Instrument Action, and press
.
8.
Ensure that Logic State reflects “Open,” change by pressing
.
Pneumatic Connections
80
i
Input/Output Configuration
Содержание Mercury Freedom 80i
Страница 1: ...Mercury System Manual Part Number 105933 00 28Jan2008 ...
Страница 6: ......
Страница 10: ......
Страница 12: ......
Страница 14: ......
Страница 42: ......
Страница 66: ......
Страница 79: ...Optional Equipment Nitrogen Generator Thermo Fisher Scientific Mercury System Manual 6 9 ...
Страница 100: ......