Introduction
Model 82i
Thermo Fisher Scientific
Mercury System Manual
1-5
Clean, dry, pressurized (~85 psig) air feeds two electo/pneumatic pressure
transducers which adjust and maintain output pressure to the Model 83
i
probe eductor and blowback pneumatics (
Figure 1–3
).
Pressurized (~85 psig) Hg-free zero air or nitrogen feeds a third
electo/pneumatic pressure transducer which adjusts and maintains dilution
gas pressure to the Model 83
i
probe dilution module.
Three electronic pressure transducers (
Figure 1–3
) associated with the
individual regulators are also contained within the Model 82
i
, as well as an
electronic vacuum transducer monitoring the dilution module vacuum in
the Model 83
i
.
Figure 1–3.
Model 82
i
Probe Controller Flow Diagram
The Model 82
i
communicates with the Model 80
i
via a RS-485 link. The
Model 80
i
acts as the master component in the Mercury Freedom
™
System with both the Models 81
i
and 82
i
slaved to the analyzer. All
commands are initiated by the Model 80
i
. The Model 82
i
receives
commands and sends information to the analyzer.
The Model 82
i
also receives two 4-20 mA signals from the probe critical
orifice pressure transducer and the fast loop venturi pressure transducer.
Pneumatics
Electronics
Содержание Mercury Freedom 80i
Страница 1: ...Mercury System Manual Part Number 105933 00 28Jan2008 ...
Страница 6: ......
Страница 10: ......
Страница 12: ......
Страница 14: ......
Страница 42: ......
Страница 66: ......
Страница 79: ...Optional Equipment Nitrogen Generator Thermo Fisher Scientific Mercury System Manual 6 9 ...
Страница 100: ......