Operating Environment
Air Conditioning
3-6
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
Thermo Scientific
Air Conditioning
The air conditioning system must be capable of maintaining a constant
temperature as specified in
in the immediate
vicinity of the system.
Thermo Fisher Scientific recommends the installation of an air
conditioner if the specified limits are exceeded due to unfavorable
climatic conditions. For reliable operation of the iCAP Qnova Series
instrument, a stability of the room temperature < 2 °C/h is required.
Humidity
The relative humidity of the operating environment must be between
20 and 80%, with non-condensing and non-corrosive atmosphere. It is
recommended that your laboratory be equipped with a temperature and
humidity monitor to ensure that your laboratory is always within the
required temperature and humidity specifications.
Altitude
The iCAP Qnova Series instruments are designed for indoor use at an
altitude of up to 2000 m (6500 ft) above sea level. For altitudes above
2000 m, contact Thermo Fisher Scientific.
NOTICE
Do not locate the instrument under an air duct, near windows, or near
heating and cooling sources. Temperature fluctuation of 1 °C or more
over a 10 minutes period can affect performance.
NOTICE
Operating an iCAP Qnova Series instrument at very low humidity
might cause the accumulation and discharge of static electricity, which
can shorten the life of electronics components. Operating the system
at high humidity might cause condensation, oxidation, and short
circuits.