Consumables
Gas Supply iCAP TQ/TQs ICP-MS
Thermo Scientific
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision A)
5-21
Pure Hydrogen or Mixtures with Helium (for QCell)
See
for details of the hydrogen gas requirements.
Oxygen or Mixtures with Helium (for QCell)
See
for details of the oxygen gas requirements. All mixture
ratios are permitted.
Supply Rate
max. 12 mL/min
Pressure min.
0.05 MPa
(0.5 bar)
max. 0.15 MPa (1.5 bar)
using a two-stage pressure
regulator (200 bar to 3 bar)
Table 5-18.
Gas requirements for argon (for QCell)
Gas Requirement
Specification
Table 5-19.
Gas requirements for hydrogen (for QCell)
Gas Requirement
Specification
Purity
99.999%
Maximum Water Content
2 ppmv
Supply Rate
max. 15 mL/min
Pressure min.
0.05 MPa
(0.5 bar)
max. 0.15 MPa (1.5 bar)
using a two-stage pressure
regulator (200 bar to 3 bar)
Table 5-20.
Gas requirements for oxygen (for QCell)
Gas Requirement
Specification
Purity
99.999% or better
Maximum Water Content
2 ppmv
Supply Rate
max. 2 mL/min for pure gas
Pressure min.
0.05 MPa
(0.5 bar)
max. 0.15 MPa (1.5 bar)
using a two-stage pressure
regulator (200 bar to 3 bar)