Operating Environment
Heat Generation and Dissipation of System
Thermo Scientific
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
3-5
Heat Generation and Dissipation of System
Heat generation and heat dissipation of the system depend on the
equipment employed. However, the iCAP Qnova Series system must
always be operated with active plasma and instrument exhaust. For a
description of the exhaust system, see
For estimated values for the average heat dissipation of the mass
spectrometer and other heat sources during analysis, see
Table 3-3.
Heat generation for a typical instrument
Module
Heat Generation [W]
at 50 Hz
Heat Generation [W]
at 60 Hz
Instrument with plasma
and heat exhaust port
connected
350
350
Fore Vacuum Pump
900
1100
Recirculating Chiller
2650
2650
Monitor
25
25
Computer
75
75
NOTICE
Water-cooled chillers minimize the heat dissipation into the laboratory
environment. However, a constant source of running water must be
available for this optional chiller.
NOTICE
The values listed in
are for a laboratory temperature of
23 °C and a cooling water temperature of 20 °C. A higher room
temperature and/or lower cooling water temperature will increase the
heat generation.