Operating Environment
Exhaust System
3-4
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
Thermo Scientific
A purge connection is provided in the Y-piece of the exhaust line, see
. This connection can be used when purging the CR gas
.
Figure 3-2.
Exhaust ports joined with Y-piece (option)
Figure 3-3.
Y-piece with capillary at purge connection
Table 3-2.
Specifications for the exhaust
Exhaust type
Hose ID [mm]
Exhaust Velocity [m/s]
Exhaust Flow [m
3
/h]
Plasma
63
6 to 8
67 to 90
Instrument
heat
63
4 to 6
45 to 67
Joined
exhaust
a
a
optional
100
5 to 7
150 to 200
purge connection
NOTICE
The exhaust system must be tested for leakage before you connect it to
the instrument.