Line Power
Electrical Power Requirements
4-6
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
Thermo Scientific
Power cables and connectors for the options are standard equipment
delivered by the manufacturers.
For connecting additional devices such as recirculating chillers, see
BRE0006802
iCAP Qnova CountKit
China
(China)
CH2-16P
YP-03
2+2
BRE0006804
iCAP Qnova CountKit TH_Type O_TH1
Thailand
(Thailand)
TH1-16P
2
Table 4-2.
Country kits cables iCAP Qnova Series, continued
P/N
Designation
Socket for MS &
fore vacuum pump
Socket for PC &
monitor
Qty.
CAUTION
High Leakage Current.
Additional Grounding Required.
For safe operation of the instrument and the quality of your
measurement, an additional grounding path between the instrument
and the fore vacuum pump must be established. The green/yellow
grounding cable delivered with the control cable for the fore vacuum
pump must be connected to the grounding bolt on the instrument (see
) and the grounding screw on the fore vacuum pump (see
and
respectively). If you need a longer
grounding cable, contact Thermo Fisher Scientific. The mounting
instructions delivered with the grounding cable must be strictly
observed. Contact your certified Thermo Fisher Scientific service
representative for more information.