Line Power
Electrical Power Requirements
4-2
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
Thermo Scientific
Electrical Power Requirements
The iCAP Qnova Series instruments are designed to operate at a
nominal voltage of 200-240 V, 50/60 Hz. For details, see
The iCAP Qnova Series instruments must have a separate “clean” line
leading to a main fuse to guarantee disturbance-free operation.
Power Connections
The electrical wall outlet for the main power of the iCAP Qnova Series
instruments should be located at the wall near the intended location of
the instrument, ideally at the left side of the instrument.
The instrument is equipped with a standard IEC 60320 C20 socket and
will be shipped with power cables suitable for you country. Depending
on your area, different types of plugs and cables may be required, see
Table 4-1.
Basic power requirements
Specification
Value
Nominal Voltage
200-240 V AC, 50/60 Hz AC, single phase
Working Voltage
180-264 V AC
Power
apparent power: 3000 VA;
effective power: 2200 W
Wire
Single-phase, 2-pole, 3-wire grounding
Fuse
Thermo Fisher Scientific recommends fusing
the wall receptacle with 15/16 A (Europe) and
20 A (North America, Japan).
NOTICE
Customers in North America and Japan
Systems installed in areas
with low line voltage (208 V in North America, 200 V in Japan),
experience voltage sags during high use periods that might place the
line voltage below the operating parameters discussed in this section.
In this case, you must protect your instrument by using a suitable
power conditioner or uninterruptible power supply (UPS).
NOTICE
When using a boost transformer or power conditioning device with
higher output than input voltage, the input voltage of the instrument
must not exceed 264 V, even when line voltage fluctuations of ±10%
occur at the input of the transformer/power conditioning device.