Site Preparation
Dimensions and Weights
Thermo Scientific
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
2-3
shows the external dimensions of the system. All values are
given in millimeter.
Ensure that the Thermo Fisher Scientific field service engineer has access
to the left and right side of the instrument.
Moving the Instrument
Figure 2-1.
Dimensions of iCAP RQ ICP-MS
1102
1020
665
772
Front View
Left View
Right View
Top View
946
665
748
136
63
NOTICE
After arrival, the instrument must be stored as packed by Thermo
Fisher Scientific for at least 24 hours at an environmental condition of
15 to 35 °C and a relative humidity of 20 to 80% before connecting it
to the power supply.