Consumables
Permitted Materials
Thermo Scientific
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision A)
5-27
Permitted Materials
The iCAP Qnova Series ICP-MS instrument is designed to be operated
with the materials listed in the following tables.
•
: Materials and their use: applicable for all instruments
•
: Gases for CCT use for iCAP RQ ICP-MS
•
: Gases for CR use for iCAP TQ ICP-MS and iCAP TQs
ICP-MS
•
: Liquids and sample matrices applicable for all
instruments
list the permitted gases used for CCT in the iCAP RQ
ICP-MS instruments.
Table 5-27.
Materials and their use
Material
Used for
Plasma Gas
Argon
Used to maintain the plasma and to actuate pneumatic
controls.
Additional Gases:
Argon
Used for gas dilution of high matrix samples. A small amount
of methane (CH
4
) might be added for optional sensitivity
enhancement for arsenic (As) and selenium (Se).
NOTICE
Maximum content of methane: 4.5%.
Nitrogen or mixtures with helium and/or
argon
Used to enhance sensitivity.
Oxygen or mixtures with helium and/or
argon
Used for the analysis of organic matrices.
Table 5-28.
iCAP RQ ICP-MS: Gases for CCT use
Material
Used for
Helium
Used for the removal of polyatomic interferences by kinetic
energy discrimination.
Oxygen or mixtures with helium
Used for the removal of polyatomic and isobaric interferences
by mass-shift reactions (oxidation).
iCAP RQ ICP-MS 1-Channel Systems
a
:
Hydrogen mixtures with helium
Used for the removal of polyatomic and isobaric interferences
by charge-transfer reactions and mass-shift reactions
(hydrogen adducts).
NOTICE
Maximum content of hydrogen: 4.5%.