Operating Environment
Temperature
3-2
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
Thermo Scientific
Temperature
The mass spectrometer is designed to operate at the temperatures
specified in
Table 3-1.
Temperature requirements for iCAP Qnova Series instruments
Specification
Value
Laboratory Temperature
15 to 35 °C
Optimum Operation Temperature
18 to 21 °C
NOTICE
As the laboratory temperature increases, system reliability decreases.
All electronics components generate heat while operating.