UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Operating Instructions
Page
205
7.
Observe the following:
♦
If the module was powered off by the power switch for a longer period (≥
1 day),
allow a warm-up period of 24 hours to achieve optimum reproducibility.
♦
If the module was powered off by the power switch less than one day, allow a warm-
up period of 60 minutes to achieve full accuracy.
7.7.3 Replacing the Filter Frit in a ProFlow Flow Meter
This section applies for ProFlow flow meters only.
The flow meter inlets on the ProFlow flow meter accommodate a built-in inline filter each.
If a filter frit is clogged, replace the filter frit.
Description
Part No.
Filter frits for inline filter (2 frits, porosity: 2 µm)
6268.0036
To replace the filter frit in an inline filter on the inlet of the ProFlow flow meter, follow
these steps:
1.
Stop the pump flow.
2.
On the ProFlow flow meter, disconnect the capillary from the inlet.
3.
Open the inline filter with a wrench (size 13 mm). Remove the filter frit.
4.
Pipette a drop of solvent into the frit holder and insert the replacement frit. Close the
filter housing. The solvent keeps the frit in the holder during closing.
5.
Perform a
Detailed Leak Test
for the flow meter (
→
page 134).
Содержание Dionex UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Страница 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Страница 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Страница 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Страница 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Страница 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...